TEROS 11 | Volumetric Soil Moisture Sensor + Temperature | METER

The TEROS 11/12 sensors are designed to be installed in mineral soils, many types of growing media, and other porous materials. This manual guides the customer through the sensor features and describes how to use the sensor successfully. TEROS 11/12 sensors are accurate tools for monitoring volumetric water content (VWC), temperature in soil and soilless substrates, and EC (TEROS 12 only). The TEROS 11/12 determines VWC using capacitance/frequency-domain technology. The sensor uses a 70-MHz frequency, which minimizes textural and salinity effects, making the TEROS 11/12 accurate in most mineral soils. The TEROS 11/12 uses a thermistor to measure temperature and EC (TEROS 12 only) using a stainless-steel electrode array. Prior to use, verify the TEROS 11/12 sensors arrived in good condition. METER recommends testing the sensors with the data logging device and software before going to the field.


Specification

Measurement Specification

Volumetric Water Content (VWC)

Range

Mineral Soil Calibration0.00–0.70 m³/m³
Soilless Media Calibration0.0–1 m³/m³
Apparent dielectric permittivity (εa)1 (air) to 80 (water)
NOTE: The VWC range is dependent on the media the sensor is calibrated to.
A custom calibration will accommodate the necessary ranges for most substrates.
Resolution0.001 m³/m³

Accuracy

Mineral Soil Calibration±0.03 m³/m³ typical in mineral soils that have solution EC <8 dS/m
Medium Specific Calibration±0.01–0.02 m³/m³ in any porous medium
Apparent dielectric permittivity (εa)1–40 (soil range) , ±1 εa (unitless)
40–80, 15% of measurement

Dielectric Measurement Frequency

70MHz


Temperature

TERROS 12

Range−40 to +60 °C
Resolution0.1 °C
Accuracy±0.5 °C from −40 to 0 °C
±0.3 °C from 0 to +60 °C

NOTE: Temperature measurement for applicable sensors may not be accurate if sensor is not fully immersed in the medium of interest due to longer equilibration time


Communication Specifications

Output

DDI serial or SDI-12 communications protocol


Data Logger Compatibility

Data acquisition systems capabl

METER ZL6, EM60, and Em50 data loggers or any data acquisition system capable of 4.0- to 15-VDC power and serial or SDI-12 communication


Physical Specifications

Dimensions

Length9.4 cm (3.70 in)
Width2.4 cm (0.95 in)
Height7.5 cm (2.95 in)

Needle Length

5.5 cm (2.17 in)


Operating Temperature Range

Minimum–40 °C
TypicalNA
Maximum+60 °C

NOTE: Sensors may be used at higher temperatures under certain conditions; contact Customer Support for assistance.


Cable Length

5 m (standard)
75 m (maximum custom cable length)

NOTE: Contact Customer Support if a nonstandard cable length is needed.


Cable Diameter

0.165 ± 0.004 (4.20 ± 0.10 mm) with minimum jacket of 0.030 (0.76 mm)


Connector Types

Stereo plug connector or stripped and tinned wires


Stereo Plug Connector Diameter

3.50 mm


Conductor Gauge

22 AWG/24 AWG drain wire


Electrical and Timing Characteristics

Supply Voltage (VIN to GND)

Minimum4.0 VDC
TypicalNA
Maximum15.0 VDC

Digital Input Voltage (logic high)

Minimum2.8 V
Typical3.6 V
Maximum5.0 V

Digital Input Voltage (logic low)

Minimum-0.3 V
Typical0.0 V
Maximum0.8 V

Digital Output Voltage (logic high)

MinimumNA
Typical3.6 V
MaximumNA

Power Line Slew Rate

Minimum1.0 V/ms
TypicalNA
MaximumNA

Current Drain (during 25-ms measurement)

Minimum3.0 mA
Typical3.6 mA
Maximum16.0 mA

Current Drain (while asleep)

MinimumNA
Typical0.03 mA
MaximumNA

Power-Up Time (DDI serial)

Minimum80 ms
TypicalNA
Maximum100 ms

Power Up Time (SDI-12, DDI disabled)

Minimum145 ms
Typical155 ms
Maximum165 ms

Measurement Duration

Minimum25 ms
TypicalNA
Maximum150 ms

Power-Up Time (SDI-12)

MinimumNA
Typical245 ms
MaximumNA


COMPLIANCE

Manufactured under ISO 9001:2015
EM ISO/IEC 17050:2010 (CE Mark)
2014/30/EU and 2011/65/EU
EN61326-1:2013 and EN55022/CISPR 22